Analyzer AMI Soditrace

Specifications

Sodium measurementSodium glass electrode, screw cap.
Calomel reference electrode, screw cap
Measuring range0.001 ppb ... 10 ppm
Accuracy±0.005 ppb or ±10 % of reading
Reproducibility±0.001 ppb or ±5 % of reading
Response time120 s (90%)
Automatic-
  • 3 point calibration
  • electrode regeneration
  • temperature compensation
  • pH monitoring and control

 

Power supply

 

 

Voltage100-240 VAC (±10%), 50/60Hz
or
10-36 VDC
Power consumptionmax. 35 VA

 

Sample conditions

Flow ratemin. 100 ml/min.
Temperature5 ... 45 °C (41 ... 113 °F)
Inlet pressure0.3 ... 3 bar (4 ... 43 PSI)
Outlet pressureambient pressure
pH Value≥ pH 7.0
Ammonium concentration< 10 ppm
Acidity< 50 ppm (CaCO3)
Suspended solids< 10 ppm, no oil, no grease

 

For further information see data sheet.

IXA2461X0X0_AMI_Soditrace