Flow cell QV-Flow SS316L L70, 1m

Specifications

  • Flow cell made of stainless steel SS316L.
  • 1 insert for one sensor with ¾" NPT thread, fitting length max: 40 mm.
  • Sample temperature: 0 .... 60 °C
  • Pressure: max. 2 bar at 60 °C
    The sample outlet has to be pressure-free against atmosphere.
  • Sample flow: 3 .... 25 l/h
  • Process connection:
    - Inlet: ¼" Swagelok
    - Outlet: SERTO angle + 6 mm tube
  • Length: without Swagelok: 91 mm
    with Swagelok + sensor (with cable): ca. 175 mm
  • Width: ca. 140 mm
  • Front-to back size (mounted): 75 mm

IXA8343515X_QV_Flow_70