Swansensor Shurecon S

Specifications

Recommended conductivity range0.1 µS/cm ... 100 mS/cm
Accuracy± 1% or ± 0.2 µS/cm; whichever is greater
Cell constant k0.45 cm-1
Temperature sensor typePt1000, DIN class A

Operating conditions:
- max. temperature120°C at 6.5 bar
- max. pressure12 bar at 20°C 
- sterilizable at120°C / 5 bar / 30min.

Materials:
- BodyPPSU
- Shaftstainless steel SS316L
- Current electrodesSS 316L
- IsolationPEEK
- Voltage electrodesPlatinum
- CoverPVDF

Electrical connection

Plug M16 male (IP67)
Process connectionG1" thread
Free space around sensor tip20 mm

IXA87341000_Sensor_Shurecon_S