DISCONTINUED: Chematest 20N

Specifications

Scope of supply:
Photometer in carrying case (38x27x10 cm) with 4 glass cuvettes, 1 cuvette lid, 2 glass rods, 1 ball pipette, 3 sample flasks, 1 waste flask, 1 note pad, 1 waterproof pen and 4 batteries.


Instrument:

  • Photometer, microprocessor controlled
  • Digital display
  • Battery powered with economical power circuit for about 2000 determinations

Dimensions, weight:
  • Instrument: 10 x 20 x 4 cm; 450 g
  • Carrying case: 32 x 25 x 5 cm; 1.200 kg (complete)

Temperature measurement (Option):
  • Electronic thermometer
  • Measuring range: -50 to +170°C
  • Resolution: 1°C
  • Accuracy: ± 1°C
  • Probe: InAisi316, length 130 mm, Ø 3 mm
  • Size: 142 x 29 x 15 mm
  • Battery: 4 x 1.4 V
  • Battery life: approx. 1000 hours

Photometric specifications:

Parameter

Measuring range mg/l

lowmediumhigh
Chlorine0.00- 2.49- 4.9- 10
Chlorine-dioxide0.00- 4.99- 9.9- 20
Ozone0.000- 0.499- 0.99- 2.5
Iodine0.00- 9.99- 19.9- 35
Bromine0.00- 4.99- 9.9- 20
Accuracy:
± 1%± 2.5%± 5 %


of measuring range


Parameter

Measuring range

Accuracy
(of measuring range)
pH-value6.50- 8.00 pH± 0.1 pH
Cyanuric acid0- 100 mg/l± 10%
Iron0.00- 2.5 mg/l± 1%
Aluminium0- 0.8 mg/l± 5%

  • Repeatability better than accuracy
  • Automatic range switching