Analyzer AMI Codes-II CC AC

Specifications

Disinfectant measurementMeasuring range [ppm]Accuracy
Free Chlorine, monochlorine, total residual chlorine0.001.00 ±0.01 ppm
1.003.00±0.06 ppm
3.005.00±0.2   ppm
Combined chlorineCalculated difference of total residual chlorine and free chlorine.
Measurement time2 min.
Cycle time free chlorine1 ... 12 min.
Cycle time total residual chlorine6 ... 60 min.
pH measuring range2 ... 12
Temperature measuring range-30 ... 100 °C

 

Transmitter Specifications

Power Supply100-240 VAC (±10%), 50/60Hz or 10-36 VDC
Power consumptionmax. 35 VA

 

Sample conditions

Water consumptionmin. 10 l/h
Water inlet pressure0.15 ... 2 bar
Sample temperature5 ... 50 °C

 

Panel

Dimension400 x 850 x 200 mm
MaterialPVC
Weight9.0 kg

IXA2544X7000 AMI_Codes-II_CC