A-89.811.060
Set of fuses for AMI
Fuse set for various instruments
Details
4 fuses 1 AT 250V for relay outputs and for AMI =2.3-C power,
1 Fuse 1.6 AT 250V for AMI 2.4, 2.5 power and AMI 2.6 AC power,
3 fuses 1AF 125V for signal outputs,
1 fuse 3.15 AT 250V for AMI 2.5 DC power,
1 Fuse 1.25 AF 250V for AMI INSPECTOR V2 (AMI AKKU),
4 fuses 12.5 AT 250V for AMI DEGASSER.
The fuse set for AMI transmitters consist of following fuses:
4 pcs 1 A, slow
1 pcs 1.6 A, slow
1 pcs 3.15 A, slow
3 pcs 1 A, fast
4 pcs 12.5 A, slow
1 pcs 1.25 A, fast
Use for following:
the 12,5 A fuses are for the Degasser controller board only, whereas the others are for the AMI Mainboard.
Relay 1, Relay 3 and Alarm relay of the AMI Mainboard require the 1 A, slow fuses
Signal Output 1 and Signal Output 2 of the AMI Mainboard require the 1 A, fast fuses
The instrument Power supply fuse is 1 A, slow for AMI Mainboards up to version 2.3-C.
For AMI Mainboards versions 2.4, 2.5 and 2.6 AC the fuse for the power supply is 1,6 A, slow.
For AMI Mainboards version 2.6 DC the fuse for the power supply is 3.15 A, slow.
The AMI Inspector Version 2 requires the 1,25 A, fast fuse