A-23.531.100
Monitor AMI-II Rescon AC
Process measuring device for the conductivity / specific resistance of ultrapure water. Complete system with transmitter, Swansensor RC U and flow cell QV-HFlow L130 with sample flow measurement on a stainless steel mounting plate. 100-240 VAC, 50/60 Hz
Product Highlights
Flow cell with integrated sensor for reliable flow measurement.
Automatic temperature compensation for different sample conditions.
Automatic measurement range switching for optimal performance.
Factory tested and calibrated on high-purity water for reliable operation.
Stainless steel flow cell with needle valve and digital sample flow measurement.
Details
The AMI-II Rescon is a complete system for the continuous, automatic measurement of specific resistivity and conductivity in high-purity water. Designed for critical applications such as the semiconductor industry. The system offers real-time data, allowing for precise control of water quality and compliance with strict industry standards, ensuring operational efficiency and product integrity.
Process instrument for the measurement of conductivity / specific resistance of ultrapure water.
Complete system with transmitter/controller AMI-II Rescon, Swansensor RC U and flow cell QV-HFlow L130 with sample flow measurement on 280 x 850 mm stainless steel mounting plate.
| Conductivity Measurement | |
|---|---|
| Measuring range | 0.01 to 18.24 MΩ×cm |
| Measuring range | 0.055 to 1000 μS/cm |
| Accuracy | 0.01 to 18.24 MΩ×cm ±0.5 % |
| Accuracy | 0.055 to 20.00 μS/cm ±0.5 % |
| Accuracy | 20.00 to 1000 μS/cm ±1 % |
| Measured value is compensated to | 25 °C |
| Temperature Measurement | |
| Measuring range | −30 to +130 °C |
| Accuracy (0-50 °C) | ±0.2 °C |
| Resolution | 0.1 °C |
| Temperature compensation | non-linear for high purity water, linear with coefficient, various chemicals |
| Sample conditions | |
| Flow rate | 20 to 100 L/h |
| Temperature | 0 to 50 °C |
| Inlet pressure | max. 2 bar |
| Outlet pressure | pressure free |
| Power Suppy | |
| AC Version | 100 - 240 VAC (± 10 %), 50/60 Hz (± 5 %) |
| DC Version | 10 - 36 VDC |
| Power consumption | max. 35 VA |