A-87.342.150
Swansensor Shurecon P
4-Electrode conductivity sensor with built-in Pt1000-temperature probe. Excellent stability and fouling resistance. For the connection to the Solicon4 conductivity transmitters.
Details
Sensor for the measurement of specific conductivity.
For applications in surface water, potable water and cooling water.
Four electrode design with integrated temperature sensor.
Unaffected by fouling.
No measuring errors due to polarization effects.
Insertion depth only 30mm
Sensor with directly attached cable.
For the use together with SWAN conductivity transmitters AMI Solicon4 or AMU Solicon4.
| Recommended conductivity range 0.1 µS/cm ... 100 mS/cm | |
|---|---|
| Accuracy | ± 1.5% or ± 0.2 µS/cm; whichever is greater |
| Cell constant k | ~0.4 cm-1 |
| Temperature sensor type | Pt1000, DIN class A |
| Operating conditions: | |
| normal operating temperature | ≤ 50°C |
| max. temperature short-time | 90°C; at ambient pressure |
| max. pressure | 10 bar at 25°C |
| Materials: | |
| Shaft end tip | PVDF |
| Electrodes | Platinum |
| Protection class | IP 68 |
| Process connection | on Ø16mm sensor head or Ø12mm shaft |
| Electrical connection | Sensor with integrated cable with end sleeves |