A-13.542.100
Transmitter AMI-II CACE
Low-maintenance monitoring of conductivity, pH, and ammonia with continuous resin regeneration for reliable water quality.
Product Highlights
Integrated calculation of pH and alkalizing agent concentration
Continuous measurement of sample flow and sample temperature for data validation
Compact dimensions and separate installation of transmitter and CACE module with 5 or 15 m cable.
Rugged IP66 aluminum housing
Fast response time after start-up.
Factory tested – Ready for installation and operation.
Details
The AMI-II CACE, combined with the SWAN CACE Module, provides low-maintenance online monitoring of specific conductivity and cation conductivity with continuous EDI resin regeneration. It also calculates sample pH and alkalizing-agent levels such as ammonia. The system ensures uninterrupted steam and condensate quality monitoring without regular resin exchange, rinsing cycles, or the risk of resin exhaustion—ideal for power and industrial water applications.
Sensors
For use with SWAN CACE Modules:
Connections for two Swansensors UP-CON1000 with built-in Pt1000 temperature sensor (specific conductivity and conductivity after cation exchange), EDI Module and for a digital sample flow meter.
User Interface
Big backlit LC display for measured values, sample temperatures, sample flow and operating status.
Easy user menus in English, German, French and Spanish. Simple programming of all parameters by keypad.
Multi-level user management for controlled access to settings and other user menus.
SD-card included for storage of measuring data records at a selectable interval, and for export of history data (events and results of instrument routines).
Connections
2 signal outputs (0/4 - 20 mA) for measured values or continuous control.
2 potential-free contacts programmable as limit switches,controllers or timers (contact NO).
1 input for potential-free contact for hold or remote-off function.
All in- and outputs protected against overvoltages.
Communication Options
2 additional signal outputs (0/4 - 20 mA), galvanically separated.
RS485 interface with Modbus RTU or Profibus DP protocol, galvanically sepa-rated.
HART interface.
| Conductivity Measurement | |
|---|---|
| Measuring Range | 0.055 to 1000 µS/cm |
| Resolution | 0.001 µS/cm |
| Accuracy (at 25°C) | ±1 % of measured value or ±1 digit (whichever is greater). |
| Response time (t90, specific conductivity) | < 5 sec. |
| pH and Alkalizing Reagent Calculation | |
| Range pH (25 °C) | 7.5 to 11.5 |
| Range concentration (e.g. ammonia) | 0.01 to 10 ppm |
| Power Supply | 100 - 240 VAC (± 10 %) / 50/60 Hz (± 5 %) / 10 - 36 VDC |
| Power Consumption | max 35 VA |
| Dimensions | 180 x 142 x 94mm |