
A-15.503.100
Transmitter AMI-II Fluotrace Oil
Transmitter for the detection of Polycyclic Aromatic Hydrocarbons (PAH) / Oil in water for Swansensor Fluotrace Oil
Product Highlights
Continuous PAH and oil monitoring
Low-concentration detection limits
Automatic cleaning control
Robust design for reliable operation
Input for flow measurement
Details
The AMI-II Fluotrace Oil transmitter is designed for reliable detection of PAH and oil traces in water using the Swansensor Fluotrace Oil. It provides continuous monitoring and additionally controls the automatic cleaning solution, supporting stable operation and reduced maintenance effort.
Instrument features
Measuring and control transmitter in a rugged aluminum enclosure (IP66).
Large, backlit LC display and simple, menu-driven operation.
Various connection options: two or optionally four analog signal outputs, two limit relays, one alarm relay and one relay input.
Modbus, Profibus or HART as an option.
| Fluorometric Measurement | |
|---|---|
| PAH (Measuring range, resolution) | 0.00 – 9.99 ppb, 0.01 ppb / 10.0 – 999.9 ppb, 0.1 ppb / 1000 – 5000 ppb, 1 ppb |
| Oil (Measuring range, resolution) | 0.00 – 9.99 ppm, 0.01 ppm / 10.0 – 150.0 ppm, 0.1 ppm |
| LOD | 0.15 ppm Oil, 5 ppb PAH |
| Accuracy | ±10% of measured value +5 ppb |
| Sample Conditions | |
| Flow Rate | approx. 5 to 50 l/h |
| Temperaure | up to 40 °C |
| Inlet pressure (25 °C) | up to 1 bar |
| Outlet pressure | pressure free |
| Power Supply | |
| AC version | 100 – 240 VAC (±10 %), 50/60 Hz (±5 %) |
| DC version | 10 – 36 VDC |
| Power consumption | max. 35 VA |
| Case | |
| Dimensions | 180 × 142 × 94 mm |











