A-13.522.100
Transmitter AMI-II Rescon AC
Reliable electronic transmitter for high-purity water resistivity and conductivity monitoring.
Product Highlights
Continuous high-purity water resistivity and conductivity measurement
Rugged IP66 aluminum housing
Fast response time after start-up.
Factory tested – Ready for installation and operation.
Details
The AMI-II Rescon transmitter continuously measures specific resistivity and conductivity in high-purity water. Wall-mountable and housed in robust IP66 aluminum, it is factory tested, calibrated, and ready for integration. Uniform menu navigation and self-explanatory maintenance ensure easy operation, user-friendly service, and reliable performance in demanding environments.
Measuring and control transmitter in a rugged, powder-coated aluminium enclosure (IP 66)
Sensors
Connections for a two-electrode conductivity sensor with built-in NT5k temperature sensor like Swansensor RC U and for a digital sample flow meter.
Temperature compensation
non-linear for high purity water, linear with coefficient, various chemicals
User Interface
Big backlit LC display for measuring value, sample temperature, sample flow and operating status.
Easy user menus in English, German, French and Spanish. Simple programming of all parameters by keypad.
Multi-level user management for controlled access to settings and other user menus.
SD-card included for storage of measuring data records at a selectable interval, and for export of history data (all user/instrument generated events and results of instrument routines).
Connection
2 signal outputs (0/4 - 20 mA) for measured values or continuous control.
2 potential-free contacts programmable as limit switches or control outputs.
2 potential-free contacts for summary alarm indication (contacts NO and NC)
1 input for potential-free contact for hold or remote-off function (interrupt control).
All in- and outputs protected against overvoltages.
Communciation Options
2 additional signal outputs (0/4 - 20 mA), galvanically separated.
RS485 interface with Modbus RTU or Profibus DP protocol, galvanically separated.
HART interface.
| Conductivity Measurement | |
|---|---|
| Measuring range | 0.01 to 18.24 MΩ×cm |
| Measuring range | 0.055 to 1000 μS/cm |
| Accuracy | 0.01 to 18.24 MΩ×cm ±0.5 % |
| Accuracy | 0.055 to 20.00 μS/cm ±0.5 % |
| Accuracy | 20.00 to 1000 μS/cm ±1 % |
| Measured value is compensated to | 25 °C |
| Temperature Measurement | −30 to +130 °C |
| Resolution | 0.1 °C |
| Power Suppy | |
| AC Verison | 100 - 240 VAC (± 10 %), 50/60 Hz (± 5 %) |
| DC Version | 10 - 36 VDC |
| Power Consumption | max. 35 VA |
| Dimensions | |
| Dimensions | 180 x 142 x 94mm |